CHAPTER 3 • Technical Aspects of the First Trimester Ultrasound Examination
INTRODUCTION
Over the past two decades the detailed ultrasound examination of a fetus before the 16th
week of gestation was made possible by two important events: the widespread adoption
of first trimester risk assessment with nuchal translucency (NT) and the improvement in
ultrasound imaging with enhanced resolution and image processing. High-resolution
transabdominal and transvaginal transducers provide images of the fetus in the first
trimester with such quality that allows for detailed anatomic evaluation. In addition, the
use of sensitive color and high-definition power Doppler improved the visualization of
the fetal cardiovascular system, including small peripheral vessels. The widespread use
of three-dimensional (3D) ultrasound technology added a new approach to fetal imaging
through the acquisition, display, and post-processing of 3D volumes. The embryo can
now be imaged on ultrasound from about the sixth week of gestation and detailed
anatomic evaluation of the fetus can be performed from about the 12 weeks of gestation
onward. This chapter provides an overview of the technical aspects of ultrasound
examination in the first trimester.
TWO-DIMENSIONAL GRAY SCALE ULTRASOUND
The quality of the two-dimensional (2D) ultrasound image is dependent on several
factors including the choice of transducers, system settings (image presets), access to
the anatomic region of interest, and magnification of the target region of interest (see
Table 3.1).
Ultrasound Transducers
Ultrasound manufacturers offer a wide range of transducers to choose from. Only a few
transducers are optimally suited for imaging the first trimester pregnancy however. Most
obstetric transducers have a frequency range between 2 and 12 MHz. Transabdominal
and transvaginal transducers that are used in the first trimester of pregnancy are
discussed in detail in the following sections.••• • ••• •••
Transabdominal Transducers
Two groups of transabdominal transducers are used in obstetric scanning: transducers
with low frequency range (2 to 5 MHz), which allow for good tissue penetration of
sound and acceptable image resolution, and transducers with high frequency range (5 to
9 MHz), which allow for improved resolution but with limited tissue penetration of
sound. The authors recommend the use of high frequency range transducers in the first
trimester when available and technically feasible, as this enables a detailed anatomic
evaluation of the fetus in keeping with existing guidelines1,2 (see Chapter 1). In the first
trimester, the use of high frequency transducers provides adequate imaging, thus
allowing for optimal nuchal and intracranial translucency evaluation along with clear
visualization of fetal organs such as brain, heart, lungs, stomach, kidneys, and bladder.
The general contour of the fetus with the surrounding amniotic fluid can be imaged (Fig.
3.1A), in addition to the skeletal system to include the skull, nasal bone, ribs, spine, and
limbs (Fig. 3.1A–E). Limitations of transabdominal high frequency transducers are
encountered when the fetus is deep in the pelvis. Recently, linear transducers, that are
commonly used for soft tissue imaging in radiology, have been adapted to obstetric
imaging.3 These linear transducers are desirable because of their high resolution with
good tissue penetration of sound. Unlike the curved array transducers, the linear
transducers have ultrasound beams that are uniform throughout all tissue levels and do
not diverge in deeper tissue. We have found linear transducers to be well adapted for
first trimester ultrasound imaging and can provide detailed anatomic evaluation of the
fetus (Fig. 3.2) with comparable resolution to that of transvaginal transducers.4
Table 3.1 • Image Optimization for Two-Dimensional Ultrasound in Gray
Scale in the First Trimester
Choose a high frequency transducer when possible
Consider using linear and transvaginal high-resolution transducers
Use the other hand to gently manipulate the uterus when performing a
transvaginal ultrasound
Combine harmonic imaging, compound imaging, and speckle reduction
when possible
Narrow the image sector
Reduce the image depth
Magnify the region of interest in order to fill one-third to half of the
ultrasound image
Use one focal zone positioned at the level of the region of interest
Adapt the dynamic range to have a high or low contrast image
Adjust image resolution• Use cine loop to return back to images stored in the recorded loop for
review
Transvaginal Transducers
When ultrasound imaging is suboptimal by the transabdominal approach due to an
increased distance between the transducer and target anatomic region (Fig. 3.3A), or
when a suspected abnormality is noted, the transvaginal approach is recommended (Fig.
3.3B and C). The main advantage of the transvaginal approach is the short distance of
the ultrasound beam to the region of interest, thus allowing for the use of higher
frequency transducers with better resolution (Fig. 3.4). Transvaginal transducers have in
general a range between 5 and 12 MHz. In the experience of the authors, fetuses with
crown-rump length (CRL) greater than 65 mm are often well imaged by transabdominal
transducers, whereas fetuses between 10 and 12 weeks of gestation and embryos before
10 weeks of gestation are better imaged by the transvaginal approach. It has also been
our experience that the NT and nasal bones are imaged easily with transabdominal
transducers. Figures 3.5 and 3.6 display the fetal abdomen and face respectively with
the transabdominal curvilinear, transabdominal linear, and transvaginal transducers.
Note that the three transducers provide adequate imaging of upper abdominal structures
(Fig. 3.5), whereas the linear and transvaginal transducers provide superior imaging for
complex anatomic regions such as the facial profile (Fig. 3.6).Figure 3.1: Various planes (A–E) obtained by a transabdominal curved array
high-resolution transducer in fetuses at 12 to 13 weeks of gestation. Plane A
represents a midsagittal view of the fetus obtained for measurement of crownrump length, nuchal and intracranial translucency, and for visualization of the
nasal bone. Plane B is an axial view of the head. Plane C is a frontal facial
view. Plane D shows two lower limbs and plane E represents the four-chamber
view.Figure 3.2: Various planes (A–F) obtained by a transabdominal high-resolution
linear transducer in fetuses at 12 to 13 weeks of gestation. Compare with
Figure 3.1. Plane A represents a midsagittal view of the fetal head. Plane B is
a frontal facial view. Plane C shows the intracerebral structures. Plane D
shows a hand with digits. Planes E and F show a sagittal and coronal view of
the fetal spine respectively with fetal kidneys noted in plane F. Note the high
resolution of these images as compared to images in Figure 3.1.Figure 3.3: A: A fetus at 12 weeks of gestation scanned transabdominally with
color Doppler at the three-vessel trachea view. Note that the image displays
decreased resolution, primarily due to the long distance between the
transducer and the region of interest; upper fetal chest in this case (yellow
arrow). Note also that the fetus is deep in the pelvis and near the cervix (white
arrow). B: A transvaginal view showing that the fetus is in a transverse lie, an
ideal fetal position for a transvaginal ultrasound examination. C: A transvaginal
ultrasound in color Doppler at the three-vessel trachea view showing improved
resolution over the transabdominal approach in A. Planes B and C are obtained
in the same fetus as plane A.Figure 3.4: Various planes (A–F) obtained by a transvaginal high-resolution
transducer in fetuses at 11 to 13 weeks of gestation. Compare with Figures 3.1
and 3.2. Plane A represents a midsagittal view of the fetal head. Plane B
shows the intracerebral structures. Plane C shows a midsagittal view of the
spine. Plane D is a four-chamber view of the fetal heart. Plane E shows a fetal
hand with digits and plane F is a coronal view of the chest and abdomen
showing the fetal kidneys. Note the high resolution of these images as
compared to images in Figures 3.1 and 3.2.
Image Presets
Image presets influence the quality of the displayed image on the monitor of the
ultrasound system. The gray scale image presets should be adapted according to the
selection of the transducer. For imaging in the first trimester, we generally recommend a
high-resolution image with high line density, in combination with harmonic imaging.
Despite recommendations to the contrary for NT measurements, we recommend
compound imaging as well as speckle reduction, for imaging of fetal anatomy in the first
trimester. A wide image angle is recommended at the initial part of the ultrasound
examination in order to measure the CRL and to assess for any gross abnormalities. The
image angle however should be narrowed in order to examine selective anatomic
regions of the fetus, such as the brain or heart. A narrow angle provides a higher imagequality with good frame rate.
Technical Skills
The technical skills of the operator performing the first trimester ultrasound examination
play a critical role in the quality of images. In general, the operator performing the first
trimester ultrasound should be well versed in the second trimester examination and
should adapt its approach to early gestation. A systematic approach to the first trimester
ultrasound, as shown in Chapter 5, standardizes the examination approach and provides
consistency in image display. In contrast to ultrasound imaging in the second trimester,
the small size of the fetus and the relatively flat maternal abdomen limits the insonation
angles in early gestation. Increased mobility of the fetus in the first trimester however
commonly overcomes this obstacle as it provides various approaches to imaging within
a relatively short time frame. Asking the mother to cough or to walk around for few
minutes can often lead the fetus to move and change position. Furthermore, applying
gentle pressure with the transducer during the transabdominal ultrasound examination
may shorten the distance to the fetus and improves imaging. With the transvaginal
approach, the transducer should be inserted gently into the vaginal canal, thus making
the examination well tolerated by most women.5 Following the introduction of the
transvaginal transducer, the operator should visualize the entire uterine cavity, including
the fetus, without magnification. Following this overview, the region of interest can be
magnified to optimize imaging and to get detailed anatomic assessment. Occasionally, a
gentle manipulation of the uterus with the other hand placed on the maternal abdomen
can lead to a change in the position of the fetus and brings the region of interest into the
focus region.Figure 3.5: Axial views of the fetal abdomen at 12 weeks of gestation in three
fetuses (A–C) using three different high-resolution transducers: A—
transabdominal curved array, B—transabdominal linear, and C—transvaginal.
Note the increased resolution in planes B and C. St, stomach.
Figure 3.6: Midsagittal views in three fetuses at 12 to 13 weeks of gestation,
imaged with three different high-resolution transducers: A—transabdominal
curved array, B—transabdominal linear, and C—transvaginal. Note the increase
in resolution and tissue characterization in C as compared to A and B. Also
note that the nasal bone (arrows) has sharp borders in B and C, as compared
to blurred borders in A. When fetal malformations are suspected, the
transvaginal approach provides more detailed assessment of fetal anatomy in
early gestation.
COLOR AND PULSED DOPPLER
Color and pulsed Doppler ultrasound has been useful in the evaluation of the first
trimester pregnancy. The application of color Doppler has been shown to be helpful in
the assessment of the fetal cardiovascular system (Fig. 3.7) and in guiding placement of
pulsed Doppler for the study of fetal vasculature. It is important to note that color and
pulsed Doppler application involves higher energy than conventional gray scale imaging
and its prudent application in early gestation is recommended. Respecting the ALARA
(as low as reasonably achievable) principle (described in Chapter 2), and using color
Doppler when indicated and in a standardized fashion, allow its safe application in the
first trimester. Pulsed Doppler application across the tricuspid valve and ductus
venosus (DV) has been used to assess aneuploidy risk and to screen for congenital heartdisease. The authors however recommend the limited use of pulsed Doppler in the first
trimester to specific indications, given its increased focused energy. In our experience,
the prudent application of color Doppler selectively on few anatomic planes in the first
trimester helps to complete the assessment of fetal anatomy. Color Doppler is
especially important for the assessment of fetal cardiac anatomy in the first trimester
(Fig. 3.8).
Color Doppler Presets
The most common use of color Doppler in the first trimester is for the examination of
the fetal heart and occasionally for the visualization of umbilical arteries, the umbilical
vein, and the DV. Ideally the examiner has to be familiar with the optimization of the
ultrasound equipment in order to properly examine the heart in early gestation.6
Improper use of color Doppler of the fetal heart bears the risk of false-negative or falsepositive diagnoses. The optimum color Doppler image is a compromise between image
quality and frame rate. Optimizing the gray scale image is essential before the
application of color Doppler. Choosing the smallest color box needed for your target
anatomic region will ensure the highest frame rate possible for the ultrasound
examination. Velocity scale or pulse repetition frequency is used to determine the range
of mean velocities within the color box. For color Doppler interrogation of the cardiac
chambers and the great vessels, a high velocity range (>30 cm per second) should be
selected. For the examination of the umbilical arteries and veins, renal arteries, or other
fetal peripheral vasculature, lower velocity ranges should be selected (5 to 20 cm per
second). Table 3.2 summarizes the presets that we commonly use for color Doppler
application in the first trimester. For a more comprehensive presentation on this subject,
the readers are referred to our previous work on the optimization of the color Doppler
ultrasound examination of the fetal heart.4Figure 3.7: A: An axial plane of the fetal chest at 13 weeks with the application
of pulsed-wave Doppler on the heart to demonstrate and document cardiac
activity. The authors do not recommend this practice given the increased
energy associated with pulsed-wave Doppler. It is recommended to use Mmode or to save a gray scale movie clip for this purpose (see Chapter 2).
When color Doppler is indicated, an application of the color box over the fetus
(B) can document cardiac activity and demonstrate an intact anterior abdominal
wall (arrow) and a normal course of the ductus venosus (DV).
Figure 3.8: Images of the fetal heart at 11 to 13 weeks of gestation, examined
with color Doppler ultrasound. A: Diastolic flow from both right (RA) and left
(LA) atrium into the right (RV) and left (LV) ventricle, respectively. B: A normal
three-vessel trachea view with aorta (Ao) and pulmonary artery (PA). C: An
oblique view showing both left and right ventricular outflow tract in systole with
the crossing of Ao and PA.
Regions of Interest for Color Doppler Application
The same anatomic regions of interest examined in the second trimester can also be
applied in the first trimester. It is important to note that not all second trimester
anatomic regions have the same clinical importance or are easy to image on color
Doppler in the first trimester. We hereby present important anatomic regions for the first
trimester color Doppler application.
Table 3.2 • Image Optimization for Color Doppler Ultrasound in the First
Trimester
Fetal Heart Peripheral Vasculature
Velocity scale High LowColor gain Low High
Color filter High Low
Color persistence Middle High
Color resolution Middle High
Heart and Great Vessels
The use of color Doppler is, in our opinion, essential for the reliable assessment of the
heart and great vessels in the first trimester. The four-chamber and the three-vessel
trachea views are relatively easy to obtain on color Doppler and provide for adequate
screening for cardiac malformations in early gestation (Fig. 3.8).4,7 In pregnancies at
increased risk for congenital heart disease, obtaining additional planes such as the fivechamber view, the short-axis and aortic arch views provides for a comprehensive
evaluation of the fetal heart in the first trimester. In selective cases, the demonstration of
the pulmonary veins draining into the left atrium can be of importance as well as the
demonstration of the course of the right subclavian artery. Detailed evaluation of first
trimester normal and abnormal fetal cardiac anatomy is presented in Chapter 11.
Abdominal Vessels
The axial plane in the lower abdomen allows for the demonstration of the two umbilical
arteries surrounding the bladder, thus confirming a three-vessel umbilical cord (Figs.
3.9A and 3.10A). In the presence of a single umbilical artery, the site of the missing
artery can be documented (Fig. 3.10B). An axial plane at the level of the mid-abdomen
allows for the demonstration of the normal abdominal wall and its umbilical cord
insertion, thus ruling out abdominal wall defects (Fig. 3.9B). In the midsagittal plane of
the fetus (NT plane), color Doppler can be applied over the abdomen to visualize the
course of the umbilical vein and DV toward the heart (Figs. 3.7B and 3.11). The narrow
size and high blood flow velocities of the DV differentiate it from the umbilical vein.
This midsagittal plane can also be used to rule out agenesis of the DV or abnormal
connections of the DV. In the same midsagittal view, two arteries appear to arise from
the abdominal aorta, namely the hepatic artery superiorly and the superior mesenteric
artery inferiorly (Figs. 3.7B and 3.11). In a slightly more angulated view, the inferior
vena cava can be visualized ascending from the middle abdomen and draining into the
right atrium8,9 (Fig. 3.12). Interrupted inferior vena cava can be confirmed in this view
when suspected in left atrial isomerism. Color Doppler applied to a coronal view of the
posterior part of the abdomen demonstrates both renal arteries arising orthogonally from
the abdominal aorta toward the renal pelves (Fig. 3.13). Detailed evaluation of first
trimester normal and abnormal fetal gastrointestinal and urogenital anatomy is presented
in Chapters 12 and 13, respectively.Figure 3.9: Transverse views in color Doppler of the fetal pelvis (A) and midabdomen (B) at 13 weeks of gestation using a linear high-resolution transducer.
Note the two umbilical arteries (arrows) surrounding the bladder (Bl.) in A and
the intact abdominal wall (open arrow) in B. Compare with Figure 3.10.Figure 3.10: Transverse views in color Doppler of the fetal pelvis in two
fetuses (A and B) at 12 weeks of gestation using a curvilinear transducer. Note
in A the presence of two umbilical arteries (arrows) surrounding the bladder
(asterisk). The absence of the right umbilical artery (?) to the right of the
bladder (asterisk) is demonstrated in fetus B. L, left; R, right.
Figure 3.11: Sagittal plane of the fetal chest and abdomen in color Doppler in afetus at 11 weeks of gestation demonstrating an intact anterior abdominal wall
(open arrow) and the umbilical artery (UA) and umbilical vein (UV) at the
insertion of the umbilical cord into the abdomen. This plane also demonstrates
the normal course of the UV in the abdomen and shows the narrow ductus
venosus (DV), connecting into the heart. This is the ideal plane for Doppler
sampling of the DV in early gestation (see Fig. 3.17). From the descending
aorta (Ao) posteriorly, the hepatic artery (Hep.A) and the superior mesenteric
artery (SMA) are seen to emerge in perpendicular orientation to the aorta (see
Fig. 3.7). The inferior vena cava is not seen in this plane as its anatomic course
runs in the right abdomen.
Placenta and Umbilical Cord
The assessment of the placental attachment and course of the umbilical artery is best
demonstrated in the first trimester on color Doppler (Fig. 3.14). The presence of
marginal or velamentous cord insertion can be easily suspected in the first trimester
given that the full length of the placenta can be imaged in one view. When umbilical
cord abnormalities are suspected in the first trimester, follow-up ultrasound
examination in the second trimester is recommended to confirm such findings.
Figure 3.12: Sagittal planes (A and B) in color Doppler of the fetal abdomen in
the same fetus at 13 weeks of gestation. Note in A, the ductus venosus (DV) in
the mid-abdomen along with the descending aorta (Ao), seen posteriorly. Theinferior vena cava (IVC) is not seen due to its anatomic course in the right
abdomen. When the probe was reoriented in B, the IVC connecting to the heart
along with the descending Ao is seen and the DV is not seen. Note the blue
color in the aorta in B (flow away from transducer) due to transducer
reorientation.
Figure 3.13: Coronal plane in color Doppler of the posterior abdomen and
pelvis at 13 weeks of gestation showing the descending aorta (Ao) with the left
and right renal arteries arising from the Ao and coursing into the kidneys
(arrows).
Intracerebral Vessels
Several articles have reported on the course of the cerebral arteries and veins in thefirst trimester in normal and abnormal conditions.10–13 Figure 3.15 shows the cerebral
arteries and veins from an axial view at the base of the skull (Fig . 3.15A),
demonstrating the circle of Willis and from the midsagittal view of the fetal head (Fig.
3.15B), demonstrating the anterior cerebral and pericallosal arteries. The application of
color Doppler of the fetal head should be reserved for pregnancies at increased risk for
central nervous system abnormalities.
Regions of Interest for Pulsed Doppler
The application of pulsed (spectral) Doppler ultrasound in the first trimester is limited
to the assessment of the maternal uterine arteries for evaluation of pregnancy risk and
fetal vessels for aneuploidy risk assessment or for fetal malformations. It is important to
note that pulsed Doppler application in the first trimester is associated with a potential
risk to the fetus and should be performed when the benefit outweighs the risk (see
Chapter 2).
Uterine Arteries
Pulsed Doppler examination of the uterine arteries in the first trimester has been used to
assess uteroplacental impedance and to integrate the results in the risk profiling for
preeclampsia14 (Fig. 3.16). This examination can be performed as part of general
screening or targeted to women with a prior history of fetal growth restriction or
preeclampsia. Given that the uterine arteries are lateral to the gestational sac, pulsed
Doppler can be performed without concern of risk to the fetus.15
The Umbilical Cord
Pulsed Doppler interrogation of the umbilical cord is rarely obtained before the 15th
week of gestation. Pulsed Doppler should not be used for confirmation of cardiac
activity, as the use of M-mode or a motion clip is preferred for that purpose.
Figure 3.14: Color Doppler showing in A the placental cord insertion in a fetus
at 13 weeks of gestation, in B a free loop of umbilical cord in a fetus at 13
weeks and in C nuchal cord in a fetus at 12 weeks of gestation resulting in a
slight increase in nuchal translucency (arrow).Figure 3.15: A: An axial plane of the fetal skull base at 13 weeks of gestation
in color Doppler showing the circle of Willis with the middle cerebral artery
(MCA). B: A midsagittal plane of the fetal head in color Doppler at 12 weeks of
gestation showing the anterior cerebral artery (ACA), the proximal portion of
the pericallosal artery (Peric.A.), and the internal cerebral vein (ICV) shown
coursing posteriorly along the borders of the thalamus. Blood flow in the
sagittal sinus (SS) is also demonstrated anteriorly.Figure 3.16: Color and spectral Doppler of the uterine artery in a pregnancy at
12 weeks of gestation (GA). Uterine artery Doppler waveforms have been used
for pregnancy risk assessment in some settings. See text for details.Figure 3.17: Color and pulsed Doppler of the ductus venosus (DV) in two
fetuses A and B. Note that the Doppler sample gate is small and is placed
within the DV, with an insonation angle of less than 20 to 30 degrees. Normal
DV Doppler waveforms show the characteristic biphasic pattern with antegrade
flow during the atrial (A) contraction phase as shown in fetus A. Note the
presence of abnormal reverse flow during the atrial contraction (A) in fetus B.
The Ductus Venosus
The most common use of pulsed Doppler in the first trimester is probably related to the
examination of the DV flow velocity waveform. In normal conditions, the DV
waveforms are biphasic with low pulsatility and with antegrade flow in the diastolic
components (a-wave) throughout the cardiac cycle (Fig. 3.17A). The presence of high
pulsatility or a reverse flow of the a-wave in the first trimester (Fig. 3.17B) increases
the risk for chromosomal anomalies, cardiac defects, and the occurrence of twin-twintransfusion syndrome in monochorionic twins.16–18 There is currently no consensus on
whether DV Doppler assessment should be a screening test performed on every fetus or
reserved as a second-line assessment in mid- and high-risk fetuses.
Tricuspid Valve
Color and pulsed Doppler examination across the tricuspid valve is commonly used inthe first trimester to assess for the presence of tricuspid valve regurgitation (TR) (Fig.
3.18). The presence of TR in the first trimester (Fig. 3.18B) has been associated with
chromosomal abnormalities.19,20 In the first trimester, TR is found in less than 5% of
chromosomally normal fetuses, in more than 65% of fetuses with trisomy 21, and in
more than 30% of fetuses with trisomy 18.19 Interrogation of other cardiac valves with
color or pulsed Doppler is reserved for fetuses at risk for valve obstruction or when a
cardiac malformation is suspected.
Other Vessels
In rare situations, clinical indications arise in early gestation for the pulsed Doppler
assessment of other fetal vessels such as the hepatic (Fig. 3.19) and middle cerebral
arteries. It has been reported that high peak velocities in the hepatic artery are present in
the first trimester in fetuses at risk for trisomy 21 (Fig. 3.19B).21 Furthermore, in rare
conditions of suspected fetal anemia in early gestation, such as in pregnancies with
serologically confirmed Parvovirus B19 infection, middle cerebral artery Doppler can
be of help in assessing for the presence of anemia.
THREE-DIMENSIONAL ULTRASOUND
3D ultrasound has been associated with a view of the fetal face or the body in surface
mode, primarily for keepsake purposes. Beyond its keepsake properties, 3D ultrasound
in the first trimester can be accurately used to reconstruct planes and to visualize
structures not seen on conventional ultrasound examination. The ability of multiplanar
reconstruction of a 3D volume is important, especially during transvaginal ultrasound,
where transducer manipulation is limited and the fetus is not in an appropriate position
to directly visualize target anatomic regions. For further information on the value of 3D
ultrasound, the reader is referred to dedicated monographs and articles on this
subject.22,23Figure 3.18: Doppler velocity waveforms across the tricuspid valve at 12
weeks of gestation in a normal fetus (A) and in a fetus (B) with trisomy 21
(T21) with severe tricuspid regurgitation (arrows).Figure 3.19: Doppler velocity waveforms of the hepatic artery at 12 weeks of
gestation in a normal fetus (A) and in a fetus (B) with trisomy 21 (T21). Note
the presence of low peak systolic velocities (18 cm per second) in the normal
fetus A, as compared to high velocities (35 cm per second) in the fetus with
trisomy 21 (T21) (B).Figure 3.20: Transvaginal 3D volume of the fetal face, obtained from an
oblique plane of the face (solid arrow in B), given that a midsagittal plane of the
fetal face could not be imaged on 2D ultrasound (open arrow in A). The volume
data are displayed in the multiplanar orthogonal mode showing A, B, and C
planes. Compare with Figure 3.21, produced after manipulation of this volume.
Multiplanar Reconstruction
Given that embryos and fetuses rarely present in the first trimester in an ideal position to
visualize all of the anatomic structures on 2D ultrasound, the acquisition of static 3D
volumes with multiplanar display of reconstructed planes can be of significant help.
Using tomographic mode display of a 3D volume, the examiner is able to show in one
image several anatomic regions of the fetus. Figures 3.20 and 3.21 show examples of
reconstructed fetal profile and NT respectively out of a 3D volume. Figures 3.22 and
3.23 show an example of the fetal head in tomographic mode. Examples of tomographic
display of the fetal chest and abdomen are shown in their respective chapters (Chapters
10 and 12). The fetal spine, limbs, profile, and internal organs such as lungs, diaphragm,
and kidneys can be reconstructed in sectional planes from a 3D volume. The brain is
probably the best organ to examine starting at 7 weeks of gestation using multiplanar
mode. Brain development can be followed from early gestation and into the earlysecond trimester. Careful rotation of the volume along the X, Y and Z-axes, in
multiplanar mode, helps to display midline planes of the face (Fig. 3.22), head (Fig.
3.23), lungs, and kidneys. The use of 3D ultrasound in the assessment of fetal anatomy in
the first trimester is presented in more detail in Chapters 8 to 15 of this book.
Three-Dimensional Volume Rendering
The use of surface mode is the most commonly used 3D rendering mode in the first
trimester as it allows for optimal visualization of the developing embryo and fetus (Fig.
3.24). Images acquired using 3D surface mode of the embryo and fetus (Fig. 3.24) are
similar to images shown in embryology textbooks. As early as the 11th week of
gestation, the head, trunk, extremities, and other fetal anatomic details can be reliably
demonstrated (Figs. 3.24 to 3.27). On occasions, 3D ultrasound can better display
normal internal anatomy of the fetus in the first trimester (Fig. 3.28). Major anomalies
affecting the external surface and internal organs of the body can be well recognized in
the first trimester in 3D surface mode (Figs. 3.29 and 3.30). In the first trimester fetal
anatomy survey, the authors caution about relying on 3D ultrasound only before a
detailed evaluation of fetal anatomy is performed on 2D imaging. In addition to 2D
ultrasound examination, 3D ultrasound plays an important role in ruling out major fetal
malformations in the first trimester in pregnant women with a prior history of severe
fetal malformations. In multiple pregnancies, fetuses can be well visualized on 3D
ultrasound along with surrounding structures. The diagnosis of chorionicity in multiple
pregnancies is best performed on 2D ultrasound. (See Chapter 7 for more details on this
subject.)Figure 3.21: The result of post-processing of the volume data set displayed in
Figure 3.20. Post-processing allowed for the retrieval of the midsagittal plane in
the left upper plane (open arrow in A), with the display of the nasal bone (NB)
and nuchal translucency (NT).Figure 3.22: Transvaginal 3D volume of the fetal head at 12 weeks of
gestation displayed in the multiplanar orthogonal mode showing A, B, and C
planes. This volume was obtained from an oblique orientation of the fetal head
as shown in the upper right image with oblique falx cerebri (dashed line). Postprocessing of this volume to display important anatomic brain landmarks is
shown in Figurve 3.23 and in the surface mode display in Figure 3.28.Figure 3.23: Post-processing of the 3D volume data set shown in Figure 3.22.
Post-processing of the 3D volume included rotations and display in tomographic
mode. Five planes are shown at 2.5 mm spacing. Anatomic details of the fetal
brain that are shown in these five planes include skull ossifications (arrows), the
falx cerebri (dashed line), the choroid plexuses (CP), the lateral ventricles (LV),
the thalami (Th), the developing cerebellum (Cer), and the fourth ventricle (4V).Figure 3.24: Surface mode display of 3D volumes of three normal embryos A,
B, and C at 8, 9 and 10 weeks, respectively. Note at 8 weeks, the relatively
large size of the embryo’s head as compared to the body.Figure 3.25: Surface mode display of 3D volumes of four normal fetuses
between 11 (B) and 13 weeks of gestation (A,C,D). Note the clear display of
surface anatomy and of extremities.
Other volume rendering modes used in 3D ultrasound include the maximum mode,
which is infrequently applied in the first trimester due to the reduced level of
ossification in the fetal skeleton, the inversion mode, which is used to visualize
intracerebral ventricular system in early gestation, and the silhouette mode (Fig. 3.26C),
which has potential for more clinical applications in the future. Combining 3D with
color Doppler in glass-body mode highlights internal vasculature. This can be used in
the first trimester to visualize the fetal heart, and the arteries and veins inside the
abdomen and thorax (Fig. 3.31).Figure 3.26: Three-dimensional volume of a normal fetus at 11 weeks of
gestation displayed in surface mode and showing the effect of various postprocessing tools. The upper panel (A–C) shows the effect of augmenting the
transparency effect, with the display of fetal internal anechoic structures. In the
lower panel (D and E), adjusting light effects in D and digitally erasing
surrounding structures in E shows the fetus without a background.Figure 3.27: Three-dimensional volume in surface mode at 12 weeks of
gestation in a normal fetus (A) and in a fetus with facial dysmorphism with
abnormal ears and micrognathia (B).Figure 3.28: Transvaginal 3D volume of a fetal head at 12 weeks of gestation
displayed in the orthogonal (A,B,C) and surface display mode (3D). This is the
same volume shown in Figures 3.22 and 3.23. Rendering of the volume shows
in the right lower panel the large choroid plexuses (CP), the falx cerebri (Falx),
and the lateral ventricles (LV). Compare with Figure 3.29.Figure 3.29: Transvaginal 3D volume of the fetal head at 12 weeks of
gestation with holoprosencephaly displayed in the orthogonal (A, B, C) and
surface display mode (3D). Rendering of the volume shows in the right lower
panel fused choroid plexuses (CP), single ventricle (double arrow in A), and
absent falx cerebri. Compare with normal brain anatomy shown in Figure 3.28.Figure 3.30: Three-dimensional (3D) ultrasound images in surface mode in a
fetus at 13 weeks of gestation with a body stalk anomaly visualized from the
front (A) and back (B). The abdominal wall defect is recognized (asterisk) and
fetal deformities of body and spine are shown in panels A and B. 3D ultrasound
is optimal imaging modality in such cases as it clearly displays the extent of
fetal deformities. Compare with normal anatomy in Figure 3.25.
Figure 3.31: The left upper panel (A) shows a 3D volume of the fetal heart in
color Doppler at 12 weeks of gestation. The left lower panel (B) shows the
same volume as in A displayed in glass-body mode with transparency. The
right upper panel (C) shows a 3D volume of the fetal abdomen in high-definition
color Doppler in glass-body mode at 12 weeks of gestation. The right lower
panel (D) displays the same volume in unidirectional Doppler flow. The right
upper (C) and lower (D) panels show the spatial anatomic relationship of the
descending aorta (D.Ao), the umbilical vein (UV), the umbilical artery (UA), the
inferior vena cava (IVC), and the ductus venosus (DV). RV, right ventricle; LV
Nhận xét
Đăng nhận xét